Article 80 of Regulation (EU) 2017/745 sets the sponsor's duty to record and report adverse events during a clinical investigation, and the sentence sponsors tend to read past is the one about timing. The clock does not start when a causality assessment concludes, when a site confirms the event, or when the picture is complete enough to write a tidy report. It starts at awareness — and MDCG 2020-10/1 Rev.1 built the reporting mechanism around that fact, including an explicit allowance most safety teams never use: an initial report is permitted to be incomplete.

A narrower duty than the directives it replaced

The AIMDD and the MDD required sponsors to report essentially every serious adverse event occurring during a clinical investigation, regardless of whether the device plausibly had anything to do with it. Article 80 deliberately narrows that duty: it requires recording of every serious adverse event, but reporting is tied to a causal relationship with the investigational device, the comparator, or the investigation procedure — including where such a relationship is merely reasonably possible, not confirmed. That threshold shift matters for clinical evaluation teams building a safety reporting SOP from an old directive-era template: a procedure that still reports every SAE regardless of causality is over-reporting against the current standard, while one that waits for a finished causality assessment before reporting anything is under-reporting against the clock.

Awareness
The event that starts the Article 80 reporting clock — not causality confirmation, not site sign-off.
2 days
MDCG 2020-10/1 Rev.1's outer limit for events indicating imminent risk of death, serious injury, or serious illness.
7 days
The outer limit for other reportable events and new findings on events already reported.

Why the clock catches teams that wait to be thorough

The instinct to withhold a report until the causality assessment is finished is reasonable from a scientific standpoint and wrong from a compliance one. MDCG 2020-10/1 Rev.1 anticipates exactly this tension and resolves it in the sponsor's favor procedurally: where necessary to meet the deadline, the sponsor may submit an initial report that is incomplete, followed by a complete report once the picture firms up. A safety team that treats “we don't know the causality yet” as a reason to delay the initial report is choosing the harder, noncompliant path when an easier, compliant one is built into the guidance.

  • Serious adverse events with a causal link, confirmed or reasonably possible, to the investigational device, the comparator, or the investigation procedure.
  • Device deficiencies that might have led to a serious adverse event had appropriate action not been taken, intervention not occurred, or circumstances been less fortunate.
  • New findings on an event or deficiency already reported — the clock applies again to material updates, not only to first reports.
  • Every concerned Member State, through the national mechanisms each competent authority currently accepts, pending the EUDAMED clinical investigation module.
The report doesn't have to be finished. It has to be on time — and the guidance tells you how to do both by not asking for the same thing twice. Why the incomplete-initial-report allowance exists

Device deficiencies are not a lesser category

Teams building their CE marking strategy around Article 80 sometimes treat device deficiencies as a secondary bucket — worth logging, less urgent to report. The regulation does not support that reading. A device deficiency that might have caused a serious adverse event, if a clinician had not intervened or if circumstances had been less fortunate, sits on the same reporting clock as an event that actually occurred. That is a deliberate design choice: it closes the obvious gap a near-miss-only duty would leave, where a investigator's quick intervention would otherwise erase the reporting trigger entirely. It also means the reporting procedure belongs inside the European regulatory strategy work done before the investigation starts, not retrofitted after the first close call — Annex XV's clinical investigation plan requirements include a description of the safety reporting procedure as a named element, not an afterthought bolted onto the monitoring plan.

A safety reporting procedure that starts the clock correctly
  1. Define “awareness” in the SOP — the date a site, monitor, or sponsor employee first learns of the event, not the date causality is assessed.
  2. Build the incomplete-initial-report path in explicitly, so investigators and safety staff know it is a compliant option, not a corner being cut.
  3. Triage severity at intake, so the 2-day and 7-day tiers are applied at the point of awareness, not discovered retroactively during a deadline review.
  4. Route device deficiencies through the same triage as adverse events, rather than a separate, slower queue.

None of this is exotic once the clock's actual start point is understood. It is a procedural discipline: log awareness the moment it happens, triage severity immediately, and use the incomplete-report allowance rather than treating the deadline as negotiable while an assessment finishes. Sponsors who still run a reporting procedure built for the AIMDD's all-SAE duty, or who wait for a finished causality call before their first report, are carrying risk the current regulation and its implementing guidance were specifically written to let them avoid. Our work on post-market vigilance reporting and on substantial modification of a clinical investigation covers the adjacent clocks this one connects to; getting Article 80 right is the piece that has to hold before either of those matters.

Frequently asked questions

When does the MDR Article 80 reporting clock start?

At the sponsor's awareness of the event or of new information about one already reported — not at the point a causality assessment is completed, not at site confirmation, and not at the point the event is fully characterized. MDCG 2020-10/1 Rev.1 explicitly permits an initial report to be incomplete, with a complete report to follow, precisely so the deadline does not depend on finishing the analysis first.

What is the reporting deadline for a serious adverse event under MDR Article 80?

MDCG 2020-10/1 Rev.1 sets a tiered clock: events that indicate an imminent risk of death, serious injury, or serious illness and require prompt remedial action — including a device deficiency that could have led to one — are reportable immediately and no later than 2 calendar days after the sponsor becomes aware. Other reportable events, and new findings on events already reported, follow a 7-calendar-day outer limit. Confirm the current guidance before building an SOP around these figures.

Does every serious adverse event have to be reported under the MDR?

No. Article 80 narrows the duty to events with a causal relationship to the investigational device, the comparator, or the investigation procedure — or where such a relationship is reasonably possible. That is a deliberate departure from the AIMDD and MDD, which required reporting of all serious adverse events regardless of causality. Device deficiencies that might have led to a serious adverse event, had circumstances been less fortunate, are reportable on the same clock even without one occurring.

Sources & further reading

  1. Regulation (EU) 2017/745, Article 80 — Recording and reporting of adverse events that occur during clinical investigations. eur-lex.europa.eu
  2. MDCG 2020-10/1 Rev.1. Safety reporting in clinical investigations of medical devices under Regulation (EU) 2017/745. health.ec.europa.eu

This article is provided for general informational purposes and reflects the regulatory landscape as of November 2026. It is not legal or regulatory advice. Confirm current MDCG 2020-10/1 reporting timelines and your national competent authorities' procedures with qualified counsel before acting.