Under 21 CFR 807.87(h), every 510(k) submission must include one of two things: a 510(k) summary, described in 807.92, or a 510(k) statement, described in 807.93. FDA does not pick between them — the submitter does, at the point of filing. Most regulatory teams never weigh the decision; they default to the summary because that is what every template, every eSTAR walkthrough, and every competitor's cleared submission in the public database shows. The default is usually the right call. It is still a choice, and the statement's standing 30-day disclosure obligation is the part teams who reach for it without reading 807.93 closely tend to miss.
Two documents, two different jobs
The summary and the statement are not two ways of writing the same thing — they do structurally different work inside the submission. A 510(k) built on a defensible predicate still needs one of the two documents to close out the filing, and which one a team picks changes what happens after clearance, not just what the review division sees during it.
- The 510(k) summary (807.92). Must be written in sufficient detail to provide an understanding of the basis for FDA's substantial-equivalence determination. If the device's indications for use differ from the predicate's, the summary has to explain why those differences are not critical to the intended use. It sits in its own section of the submission, starting on a new page, ending on a page not shared with any other section, and is clearly labeled "510(k) summary."
- The 510(k) statement (807.93). Defined under section 513(i) of the FD&C Act as a statement — signed by the certifying official — asserting that all safety and effectiveness information in the submission will be made available to any person within 30 days of a written request. It excludes patient identifiers and trade-secret or confidential commercial information from what gets produced. The statement itself does not contain the underlying data; it is a commitment to produce that data later, to whoever asks.
A summary closes the disclosure question at clearance. A statement reopens it every time someone asks — and the submitter has thirty days to answer, for as long as the device is on the market. Why most submitters default to the summary
Why the summary is the default, and when the statement still makes sense
A 510(k) summary is bounded work: write it once, have it reviewed once as part of the submission, and once FDA clears the device, the summary becomes part of the public record in FDA's 510(k) database. There is no ongoing duty attached to it. A 510(k) statement defers that drafting effort, but it replaces a one-time task with a standing one — the submitter has to be able to produce the underlying safety and effectiveness data, in full, within 30 days, every time a written request arrives, for as long as the device remains on the market. That is a real operational commitment: someone has to own incoming requests, confirm what is excluded (patient identifiers, trade secrets, confidential commercial information), and assemble the rest inside the statutory window. Few submitters are set up to run that process indefinitely, which is why the summary route dominates in practice even though 807.87(h) treats both as equally valid. The statement can still make sense where a submitter has a specific reason to keep the substantial-equivalence rationale out of the public submission file at clearance — but that reason should be deliberate, not a byproduct of skipping the summary because it looked like more drafting work up front.
- Confirm the submission needs one of the two. 807.87(h) requires a summary or a statement — decide which before drafting, since the two are structured differently.
- If drafting a summary, meet the 807.92 standard. Sufficient detail to support the substantial-equivalence finding, its own labeled section, starting on a new page.
- If choosing a statement, build the 30-day process now. Someone has to own incoming requests, apply the patient-identifier and trade-secret exclusions, and produce the rest inside the window — for the life of the clearance.
- Don't default to the statement to save drafting time. A bounded, one-time summary is usually less exposure than an open-ended production duty with no end date.
None of this changes the substantial-equivalence work a 510(k) clearance strategy still has to do — the summary-or-statement choice sits downstream of predicate selection and testing, not upstream of it. But it is still a choice with real consequences, and it belongs in the same pre-submission planning conversation as labeling and the rest of what an eSTAR submission has to assemble — not a box checked after everything else is already final.
Frequently asked questions
Does every 510(k) need a 510(k) summary?
No. Under 21 CFR 807.87(h), a premarket notification must include either a 510(k) summary as described in 807.92 or a 510(k) statement as described in 807.93 — the submission needs one of the two, not specifically the summary. Most submitters choose the summary, but the statement remains a valid alternative.
What does a 510(k) statement actually commit the submitter to?
A 510(k) statement, defined under section 513(i) of the FD&C Act, is the submitter's assertion that all safety and effectiveness information in the submission will be made available to any person within 30 days of a written request, excluding patient identifiers and trade-secret or confidential commercial information. It is signed by the certifying official and does not itself contain the safety and effectiveness data — it commits to producing that data later, on request, for as long as the clearance stands.
Why do most submitters file a summary instead of a statement?
A 510(k) summary under 807.92 is written once, reviewed once, and posted to FDA's public 510(k) database once cleared — a fixed, bounded disclosure. A 510(k) statement defers that disclosure but does not eliminate it: the submitter still has to track incoming requests and produce the underlying data within 30 days, indefinitely, for as long as the device remains on the market. Most submitters treat a bounded, one-time drafting task as the lower-risk choice over an open-ended standing obligation.
Sources & further reading
- eCFR. 21 CFR 807.92 — Content and format of a 510(k) summary. ecfr.gov
- eCFR. 21 CFR 807.93 — Content and format of a 510(k) statement. ecfr.gov
- eCFR. 21 CFR 807.87 — Information required in a premarket notification submission. ecfr.gov